search
for
 About Bioline  All Journals  Testimonials  Membership  News


Journal of Cancer Research and Therapeutics
Medknow Publications on behalf of the Association of Radiation Oncologists of India (AROI)
ISSN: 0973-1482
EISSN: 0973-1482
Vol. 7, No. 3, 2011, pp. 298-303
Bioline Code: cr11075
Full paper language: English
Document type: Research Article
Document available free of charge

Journal of Cancer Research and Therapeutics, Vol. 7, No. 3, 2011, pp. 298-303

 en Additional rectal and sigmoid mucosal points and doses in high dose rate intracavitary brachytherapy for carcinoma cervix: A dosimetric study
Mahantshetty, Umesh; Tiwana, Manpreet Singh.; Jamema, Swamidas; Mishra, Sanjib; Engineer, Reena; Deshpande, Deepak & Shrivastava, Shyamkishore

Abstract

Background: Late rectal and sigmoid toxicities seen in cervical cancer patients are attributed to brachytherapy despite rectal doses within tolerance limits. The purpose of this study was to identify additional dosimetric points which may better forecast rectal complications.

Materials and Methods: Fifteen high dose rate intracavitary brachytherapy (ICA-HDR) applications with conventional X-ray and computed tomography (CT) based planning were studied. In addition to International Commission on Radiation Units and Measurement (ICRU) rectal and bladder points, proximal and distal rectal and sigmoid points were digitized on CT scans and dose volume histograms′ (DVHs′) parameters were computed and correlated.
Results: The mean ICRU, additional distal, proximal and sigmoid point doses were 486 ± 152 cGy, 527 ± 156 cGy, 401 ± 149 cGy and 838 ± 254 cGy, respectively, for a prescription of 700 cGy to point A. The mean sigmoid point dose was significantly higher than the ICRU rectal point doses (P=0.001). The high-dose sigmoid points were situated at a mean -8 mm (range -22.95 to 10.43 mm) lateral, 10 mm posterior (range -15.87 to 27.82 mm) and 31 mm (range 8.08-62.91 mm) cranial to the intracavitary applicator flange of central tandem.
Conclusions: Our dosimetric study suggests that sigmoid points and 0.1 cm 3 receive significantly higher doses than rectal points during ICA-HDR in carcinoma of the uterine cervix. No definite conclusion on reproducible spatial distribution on orthogonal X-rays could be achieved. To document and reduce sigmoid doses, some form of 3D image-based planning is necessary.

Keywords
Additional proximal and distal rectal point/s, dose volume parameters, high dose rate intracavitary brachytherapy, International Commission on Radiation Units and Measurement 38, rectal point, sigmoid point/s

 
© Copyright 2011 Journal of Cancer Research and Therapeutics.
Alternative site location: http://www.cancerjournal.net/