 en Follow-Up Results of Device Occlusion of Patent Ductus Arteriosus
Amoozgar, Hamid; Salehi, Sara; Farhadi, Pouya; Edraki, Mohammad Reza; Borzoee, Mohammad; Ajami, Gholamhossein; Cheriki, Sirous & Mohammadi, Hamid

Abstract

Background: Transcatheter patent ductus arteriosus (PDA) closure is an established procedure.
Objectives: The aim of the study was to assess midterm follow up of the Nit-Occlud coil and the amplatzer ductal occluder (ADO) closure of PDA.
Patients and Methods: In this cohort study, we collected the longitudinal data of patients who underwent percutaneous closure using coil or ADO from November 2005 to November 2013. A total of 404 patients with PDA closure by devices were included during the study period. Coil occlusion was performed in 220 patients and 184 patients underwent catheterization using ADO. Follow-up evaluations were performed with echocardiography at two weeks, two months, six months, and during the study period (in average 4.8 ± 3.8 years).
Results: The patients’ mean age was 24 months (range: 1 - 312). The catheterization was successful in 393 (97.2%) patients and unsuccessful in 11 (2.7%). Immediate complete occlusion was seen in 290 (73.7 %) patients. The occlusion rates at two weeks, two months, six months, and during the study period were 73.7%, 84%, 93.6%, 98.7%, and 99.5%, respectively. Complications occurred in 23 (5.8%) patients during or immediately after the catheterization, and device embolization with 2.7% was the most common complication. Most complications occurred in a patient with pulmonary hypertension who was less than one year old and was undergoing the first year of experience with devices.
Conclusions: Our findings showed that transcatheter occlusion of the PDA is an effective and safe intervention by coil or Amplatzer with excellent early and one-year outcomes. Pulmonary hypertension, age of less than 12 months and experience of less than one year may increase the complications of device closure.

Keywords
Transcatheter Occlusion; Patent Ductus Arteriosus; Coil Occlusion; Amplatzer Ductal Occlude; Treatment Outcome

 
© Copyright - 2016 Iranian Journal of Pediatrics
Alternative site location: http://diglib.tums.ac.ir/pub/

Home Faq Resources Email Bioline
© Bioline International, 1989 - 2024, Site last up-dated on 01-Sep-2022.
Site created and maintained by the Reference Center on Environmental Information, CRIA, Brazil
System hosted by the Google Cloud Platform, GCP, Brazil